BIOM® ready
Instructions for Use
Caution
Risk of eye damage if the working distance between the
BIOM® ready and the patient changes.
When the BIOM® ready is swung into the working position
(into the beam path), the following instructions must be
followed:
Do not use the coarse adjustment knob at the
microscope stand to adjust the height.
Do not adjust the height of the stand arm, either by
motor or manually, over the surgical area.
Do not change the patient's position by adjusting the
height of the OR- table.
Risk to the patient as a result of uncontrolled head
movements of the patient.
Take suitable measures to prevent uncontrolled head
movements of the patient during the surgical
procedure.
Avoid any contact of the front lens with the cornea.
If the lens touches the cornea, the image will become blurred immediately.
In an extreme emergency, e.g. uncontrolled downward movement of the
microscope:
Stop by pressing the Emergency Off button.
Pull / move the complete surgical microscope upwards.
Swing the BIOM® ready out of the beam path.
If contact with the eye has occurred: Swing out the BIOM® ready, or lift the bottom
part of the BIOM® ready to make it easier to clean the optics, so that you can clean
the lens with a sterile swab.
Clean the front lens with care, as it can easily break off.
Make sure that the cornea is adequately wetted with a suitable solution. This
prevents damage to the cornea and ensures optimal observation.
en
Instruction Manual
Focusing Instructions
You can manually focus the BIOM® ready with the focusing knob.
Heed the following points when focusing with the BIOM® ready:
The BIOM® ready is only needed for visualization of the posterior eye section.
After performing the surgical steps at the front section and outside of the eye, leave
the microscope at this position (height).
Before swinging it into the working position, make sure that the BIOM® ready has
been set to the shortest length.
Before beginning the focusing process, check the distance from the front lens to the
patient's eye.
Monitor the distance from the front lens to the patient's eye during the focusing
process.
Avoid the front lens from coming into contact with the patient's eye.
While focusing with the BIOM® ready, make sure that
The working height of the microscope is not adjusted.
The focusing function of the microscope is not used.