KERN KME-TM Mode D'emploi page 25

Afficheurs aux choix
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

F1.4.1 – Automatic zero setting
Automatic zero setting disabled (not available in NTEP-Mode)
OFF
Automatic zero setting within +/–0.5 d
0.5 d
Automatic zero setting within +/–1.0 d (not available in OIML-Mode)
1 d
Automatic zero setting within +/–3 d (not available in OIML-Mode)
3 d
F1.4.2 – Power up zero
Power up zero disabled
OFF
Power up zero within +/–2 %
2
Power up zero within +/–10 %
10
Power up zero within +/–20 % (not available in OIML and NTEP-Mode)
20
F1.4.3 – Pushbutton zero
Pushbutton zero disabled
0
Pushbutton zero with +/–2 % zero setting range
2
Pushbutton zero with +/–10 % zero setting range
10
(not available in OIML and NTEP-Mode)
Pushbutton zero with +/–20 % zero setting range
20
(not available in OIML and NTEP-Mode)
F1.5.1 – Automatic taring
Automatic taring enabled
On
Automatic taring disabled
OFF
F1.5.2 – Auto clear tare
Clearing tare automatically enabled
On
Clearing tare automatically disabled
OFF
F1.5.3 – Tare Interlock
The weighing platform must be unloaded to zero before the tare weight can be
On
cleared.
Function disabled
OFF
F1.5.4 – Auto tare threshold
This menu item is only available if F1.5.1=On is set.
The weighing platform must be loaded to the set value before the weight value is automatically
tared.
Factoriy setting
Possible settings
10 d
0 ... maximum load
24

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Kmn-tm

Table des Matières