Continuity Test; Capacitance Measurement - VOLTCRAFT WBM-460 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

h) Continuity test
Ensure that all objects that you wish to measure (including circuit components, circuits and com-
ponent parts) are disconnected and discharged.
- Switch the DMM on and select the
- Press the 'MODE' button to toggle the measuring function. The con-
tinuity test symbol and the Ω symbol appear on the display. Press
the button again to switch to the next measuring function.
- I nsert the red test lead into the Ω measurement socket (10) and the
black test lead into the COM measurement socket (9).
- I f the measured resistance is equal to or less than 10 Ω, the multi-
meter beeps to indicate continuity. There is no further beeping from
approx. >50 Ω. The continuity test measures resistances of up to
600 Ω.
- 'OL' (overload) indicates that the measurement range has been ex-
ceeded or that the circuit is broken.
- After taking a measurement, remove the leads from the measured
object and switch the DMM off.
i) Capacitance measurement
Ensure that all objects that you wish to measure (including circuit components, circuits and com-
ponent parts) are disconnected and discharged.
Always pay attention to the polarity when using electrolytic capacitors.
- Switch the DMM on and select the 'CAP' measuring function.
- Press the 'MODE' button three times to toggle the measuring func-
tion. The unit 'nF' appears on the display for capacitance measure-
ment. Press the button again to switch to the next measuring func-
tion.
- Insert the red lead into the V socket (10) and the black lead into the
COM socket (9).
Due to the sensitive measuring input, the display may show
a reading even with 'open' test leads. Press the 'REL' button
to measure small capacities (<600 nF). The indicator is then
set to '0'. Auto range is thereby disabled.
- Connect the two probe tips (red = positive, black = negative) to the
object that you want to measure (capacitor). The capacitance is
shown on the display after a few seconds. Wait until the display
stabilises. This may take a few seconds for capacitances greater
than 40 µF.
- 'OL' (overload) indicates that the measurement range has been exceeded.
- After taking a measurement, remove the leads from the measured object and switch the DMM off.
measuring function.
77