Alternating current A/AC True RMS
Range
600.0 µA
6000 µA
60.00 mA
600.0 mA
10.00 A
Specified measurement range: 5 - 100% of the measurement range, sine wave
Overload protection 1000 V; frequency range 50 Hz - 1 kHz
Fuse: µA/mA = ceramic high-performance fuse FF 800 mA/1000 V
10 A = Ceramic high-performance fuse FF 10 A/1000 V
≤6 A = Continuous measurement, >6 A = max. 10 s at intervals of 15 minutes
Accuracy of PEAK function: ±10% of the reading
Alternating current with clamp-on ammeter A/AC
Range
30.00 A
300.0 A
3000 A
1000 V overload protection
Resistance
Range
600.0 Ω*
6.000 KΩ
60.00 KΩ
600.0 KΩ
6.000 MΩ
60.00 MΩ
1000 V overload protection
Measurement voltage: Approx. 0.5 V, measurement current approx. 0.3 mA
*Accuracy for measurement range ≤600 Ω was calculated after deducting lead resistance from the REL function
92
Resolution
0.1 µA
1 µA
0.01 mA
0.1 mA
0.01 A
Resolution
0.01 A
0.1 A
1 A
Resolution
0.1 Ω
0.001 KΩ
0.01 KΩ
0.1 KΩ
0.001 MΩ
0.01 MΩ
Accuracy
AC
±(1.2% + 5)
±(3.0% + 20)
±(1.5% + 5)
±(3.0% + 20)
Accuracy
50 – 60 Hz
61 Hz – 1 kHz
±(0.8% + 5)
±(2.4% + 5)
Accuracy
±(0.5% + 10)
±(0.5% + 5)
±(2.5% + 10)
AC+DC