Aesculap
HiLAN pneumatic motor GA519/GA520R/GA529
Note
If coolant pump GA530 is employed for cooling/irrigating the tool blades,
connect the coolant pump between the foot control switch and the HiLAN
pneumatic motor, see TA022140.
Connecting the HiLAN pneumatic motor with hand lever to the com-
pressed-air supply
The HiLAN pneumatic motor GA529 is controlled via hand lever 17 and is
connected to the compressed-air source directly via a tube.
Coupling:
►
Make sure that slide 18 is in position "0".
►
Select the compressed-air-tube system appropriate for the systems
installed at your site.
►
Pull back unlocking sleeve (motor-side) 3 at fast-lock coupling 4 and
attach motor tube 5 to tube connection 19 on the HiLAN pneumatic
motor 1. Pull at 4 to check that the coupling is correct.
►
Join connector (source-side) 14 and connecting tube 13 to the com-
pressed-air source in the operating room.
Connecting the foot control switch to the compressed-air supply
Aesculap Dräger system
Coupling:
►
Pull back unlocking sleeve (foot control-side) 11 at connecting
tube 13 and connect the tube with fast-lock coupling 12 to connector
(foot control-side) 9 on the foot control switch.
►
Release unlocking sleeve (foot control-side) 11.
►
Insert connector (source-side) 14 of connecting tube 13 into the
Aesculap Dräger socket until it engages.
Uncoupling:
►
Pull back unlocking sleeve (foot control-side) 11 at connecting
tube 13.
►
Schnellabsperrkupplung 12 am Stecker (fußschalterseitig) 9 an der
Fußsteuerung abziehen.
►
Pull back unlocking sleeve (source-side) 15 at connector (source-
side) 14 of connecting tube 13.
►
Pull connector (source-side) 14 out of the Aesculap Dräger socket.
6
HiLAN
®
DIN connection
Coupling:
►
Pull back unlocking sleeve (foot control-side) 11 at connecting
tube 13 and connect the tube with fast-lock coupling 12 to connector
(foot control-side) 9 on the foot control switch.
►
Release unlocking sleeve (foot control-side) 11.
►
Insert the DIN connector of connecting tube 13 into the DIN socket
until it engages.
Uncoupling:
►
Pull back unlocking sleeve (foot control-side) 11 at connecting
tube 13.
►
Remove fast-lock coupling 12 from connector (foot control-side) 9 on
the foot control switch.
►
Pull back unlocking sleeve (source-side) 15 at connector (source-
side) 14 of connecting tube 13.
►
Pull connector (source-side) 14 out of the DIN socket.
5.2
Function checks
►
Always check the pneumatic motor system (HiLAN pneumatic motor,
foot control switch, tubes) by performing a trial run prior to use.
►
Make certain that the HiLAN pneumatic motor and foot control switch
are in good working order.
►
Make certain that all tubes are properly connected and free of leaks.