2.1. TipOne
Repeat Dispenser Tips
®
The minimum dispensing volume
and the maximum
1
filling capacity
is stated on each TipOne
2
Dispenser Tip (Fig. 2). The scale facilitates drawing off
partial quantities and helps you to estimate the amount
of liquid still available for operation.
The volume to be dispensed is obtained by multiplying
the set number located on the volume selection dial by
the minimum dispensing volume of the TipOne
Dispenser Tip.
Eg. 1.25 ml TipOne
®
Repeat, Minimum volume: 1=25 µl
at setting 4: 4 x 25 µl = dispensed volume: 100 µl
Better precision is achieved by using a TipOne
Dispenser Tip of small capacity with a high dial setting.
Repeat
®
2
Repeat
®
1
®
Repeat
Fig. 2
2.2. 25 ml and 50 ml
Number
Position
TipOne
of Dispensing
of Volume
Dispenser Tips
Steps
Selection Dial
48
1
Note: An adapter, included
with each RepeatOne
23
2
necessary for use with the

25 ml and 50 ml TipOne

15
3

Repeat Dispenser Tip.

11
4
Attach the adapter to the
25 ml or 50 ml TipOne
8
5
Repeat Dispenser Tip
by turning it clockwise;
The correlation between the
remove by turning it
position of the volume se-
counter clockwise.
lection dial and the number
of dispensing steps is as
shown in the table above.