Call supplier
This product complies with Directive
93/42/EEC concerning medical devices.
The launch date of this product is stated in the
CE declaration of conformity.
For more details about CE marks, refer to Typical Product
Parameters.

Symbols on Product

Unit running
Unit not running
1163146-H~07
O
Indicators
2
Symbol
O
Purity
2
O
SYSTEM OKAY
2
O
over 85%
2
O
between 73% to
2
85%
SYSTEM FAILURE
O
Below 73%
2
General
Indicator Lights (LED)
GREEN Indicator Light
YELLOW Indicator Light
A. YELLOW Solid
B. YELLOW Flashing
Sensor
Failure
Call a qualified technician
RED Indicator Light
Continuous Audible
Alarm Sieve — GARD™
Compressor Shutdown