Chauvin Arnoux SIMPLE LOGGER II Notice De Fonctionnement page 48

Masquer les pouces Voir aussi pour SIMPLE LOGGER II:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
SIMPLE LOGGER II
MODEL
Electrical
Inputs
Input connections
Measurement span according to
thermocouple
Resolution
Uncertainties(J, K, T, N, E)
Uncertainties (R, S)
Temperature coefficient
Maximum differential voltage
Sample rate
Storage rate
Storage modes
Recording length
Memory
Communication
Power source **
Battery life
Mechanical
Dimensions
Weight (with batteries)
Case
Vibration
Shock
Drop
Environmental
Operating temperature
Storage temperature
Relative humidity
Altitude
Safety
Conformity
Level of protection
*See the end of chapter
2 miniature thermocouple connectors
°F
J
- 346 to + 2192
K
- 328 to + 2501
T
- 418 to + 752
N
- 328 to + 2372
E
- 238 to 1742
R
32 to 3212
S
32 to 3212
θ < 1000°C or °F
θ ≤ - 100°C (- 148°F)
- 100°C (- 148°F)< θ ≤ + 100°C (212°F)
θ > + 100°C (212°F)
θ = 0° (32°F) to + 100°C (212°F)
θ > + 100°C (212°F)
± (0,02% of reading + 0,03°C) / °C or ± (0,02% of reading + 0,03°F) / °F for operating
temperatures from- 10°C to + 18°C and + 28°C to + 50°C
1 V (between the low points of the inputs)
8 samples maximum per recording interval
Programmable from 5 seconds to 1 day
Start/Stop, FIFO or Extended Recording Mode* (XRM™)
Programmable using DataView
240,000 measurements (512 kB). The recorded data is stored in non-volatile memory
retained even if the battery is low or removed.
2 x 1.5 V AA (LR6) cell Alkaline batteries
100 hours to >45 days (dependent on storage rate/recording length)
136 x 70 x 32 mm (5.38 x 2.75 x 1.28")
IEC 68-2-6 (1,5 mm, 10 to 55 Hz)
- 10°C to + 50°C (14°F to 122°F)
- 20°C to + 60°C (- 4°F to + 140°F)
Up to 85% to 35°C (95°F) without condensation
EN61010-1 ; 50 V CAT III ; Pollution level 2
L642
Two
J
- 210 to + 1200
K
- 200 to + 1372
T
- 250 to + 400
N
- 200 to + 1300
E
- 150 to + 950
R
S
0,1 °C or °F and θ ≥ 1000°C or °F
± [0,2% of reading + 0,6°C (1,1°F)]
± [0,15% of reading + 0,6°C (1,1°F)]
± [0,1% of reading + 0,6°C (1,1°F)]
± [0,15% of reading + 1,0°C (1,8°F)]
± [0,1% of reading + 1,0°C (1,8°F)]
®
from 15 minutes to 8 weeks
USB 2.0 optically isolated
200 g (7 oz)
UL94-V0
IEC 68-2-27 (30 G)
IEC 68-2-32 (1 m)
2000 m
IP 40
English
°C
0 to 1767
0 to 1767
1°C or °F
Page 48

Publicité

Table des Matières
loading

Table des Matières