5.2 REPLACING THE FUSE (model L111)
Using a flat blade screwdriver, press the cap of the fuse holder and unscrew a quarter turn;
When you stop pressing, the cap and fuse pop out by themselves;
Replace the fuse with an identical model: FA, 2 A / 250 V (5 x 20 mm);
Replace the cap and screw it back in a quarter turn.
5.3 CLEANING
The housing of the logger should be cleaned using a cloth moistened with soapy water. Do not put the
device into the water. Rinse in the same way. Do not use a solvent. Dry with a cloth or forced air. Dry
carefully and thoroughly before using.
For the model CL601, it is important always to keep the air gap surfaces clean. Clean these surfaces gently
with a soft cloth and oil them lightly to prevent rust. Do not use water or a solvent; leave no dirt behind.
5.4 METROLOGICAL CHECK
Like all measuring or testing devices, the instrument must be checked regularly.
This instrument should be checked at least once a year. For checks and calibrations, contact one of our
accredited metrology laboratories (information and contact details available on request), at our Chauvin
Arnoux subsidiary or the branch in your country.
5.5 REPAIRS
For all repairs before or after expiry of warranty, please return the device to your distributor.
5.6 WARRANTY
Our warranty applies, except as otherwise expressly stipulated, for twelve months counting from the date of
availability of the equipment. Extract from our General Terms of Sale, communicated on request.
The warranty does not apply following:
♦ improper use of the equipment or use with incompatible equipment;
♦ changes made to the equipment without the explicit approval of the manufacturer's engineering
department;
♦ work done on the device by a person not approved by the manufacturer;
♦ adaptation to a particular application not anticipated in the definition of the equipment or not indicated
in the user manual;
♦ damage due to shocks, falls, or flooding.
WARNING: Risk of electric shock. Remove the clamp from any conductor and
disconnect all inputs before replacing the fuse.
WARNING: Risk of electric shock. Disconnect all inputs or remove the clamp from
any conductor before cleaning.