5.3 Connect trigger

The device has the option of being externally
triggered.
C A U T I O N :

Material damage!

Do not trigger the device with signals in excess of
300,000 FPM.
N O T E :
Only use original material from the manufacturer to
connect the trigger signal.
The trigger input is isolated. The isolated input is
suitable for PNP and NPN signals. A trigger plug
suitable for this input socket is included with the
device.
1. Plug the trigger plug into input socket
INPUT (B).
2. Screw in trigger plug.
3. Assign the connection socket as per
4
2
1
DIN 41 524
(EN 60130-9)
Fig.3 – Assignment of the connection socket
1
+24 V
2
nc
3
GND
4
+Trigger
5
-Trigger
N O T E :
The device must be manually switched
between external and internal trigger signals
(
† section 6.2.2
an internal / external trigger").
.
ig. 3
5
3
– "Standard mode / How to select
EN
9