Exchanging The Cable/Rod - Balluff BMD 1L K Série Notice D'utilisation

Sonde de mesure tige et câble avec qualification sil
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Exchanging the cable/rod

Or on site by the user
electronics module.
sor from our homepage.
Under "Instrument search (serial number)" you can download the
to the sensor.
After the transfer of the sensor data, you have to verify the correct
transmission by means of a check sum. Only then, the instrument will
be ready for operation, again.
supplementary instructions "Electronics module".
out a fresh setup after exchanging the electronics or load the stored
data of the setup.
sor, you can transfer them to the replacement electronics module. A
fresh setup is then not necessary.
10.6 Exchanging the cable/rod
If necessary, the cable or rod (measuring part) of the probe can be
exchanged.
cable ø 2 and 4) or wrench size 10 (rod ø 12).
Note:
When exchanging the rod or cable, make sure that the instrument and
the new rod or cable are dry and clean.
surface, provide counterforce with another fork wrench.
the measuring rod.
3. Unscrew the loosened rod or cable manually.
4. Insert the new measuring rod carefully by hand with a screwing
5. Continue screwing in the rod manually into the opening of the pro-
6. Exert counterforce with the second fork spanner and tighten the
Rod ø 12: 10 Nm (7.37 lbf ft)
10 Diagnostics and servicing
-
-
67

Publicité

Table des Matières
loading

Table des Matières