CAUTION
Risk of infection due to moisture
Moisture encourages the growth of bacteria.
• Remove all parts from the pot or disinfector as soon as
the disinfection process is finished.
• Dry all parts.
IN BOILING WATER
• Place all the individual parts in boiling water for at least
5 minutes. Use a clean pot and drinking water.
NOTE! Risk of damage to plastic parts! Plastic will melt if it
comes into contact with the hot base of the pot. Make sure
there is enough water in the pot to prevent the individual
parts from touching the pot base.
• Shake the water out of all of the parts.
USING A STANDARD THERMAL DISINFECTOR FOR BABY
BOTTLES (NOT A MICROWAVE OVEN)
CAUTION
Risk of infection due to inadequate disinfection
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Make sure that the disinfector is clean and operating prop-
erly before every disinfection process.
• Allow the disinfection to continue until the disinfector
switches off automatically or the minimum disinfection
time stated in the instructions for use of the disinfector has
elapsed. Do not switch the device off prematurely.
Use a thermal disinfector with a runtime of at least 6 minutes.
Regarding the performance of the disinfection, the duration of
the disinfection procedure and the quantity of water required,
follow the instructions for use of the disinfector you are using.