Advanced Menu List

Display Example
P: 5.0
Heat:Off
Upper Limit
15.0 cmH20
Lower Limit
5.0 cmH20
Set Pressure
10.0 cmH20
Delay Time
45 minutes
Delay Pressure
5.0 cmH20
Tubing Length
1.8 m (6 ft)
Enable Menu
Press 
LCD Contrast Adj
50 %
Compliance Meter
123456.7 hours
Hourmeter
123456.7 hours
SE-DV54
12
aDVancED mEnu controlS
Mode
Description
This is the default screen when the SleepCube is
turned on. The actual mask pressure and heater
Auto
setting are shown. The displayed pressure may vary
All
slightly while breathing. If the heater system is not
connected, the heat setting will be blank.
This is a prescription setting which limits the
maximum pressure at which the SleepCube will
AutoAdjust
operate if in AutoAdjust mode. Your home care
Mode
provider programs this setting for you during device
setup.
This is a prescription setting which sets the
minimum pressure at which the SleepCube will
AutoAdjust
operate in AutoAdjust mode. Your home care
Mode
provider programs this setting for you during device
setup.
This is a prescription setting which controls the
CPAP
operating pressure.
Mode
Selectable delay time setting. Your provider has set
the delay time. If the unlocked symbol is shown,
you are able to adjust this setting. If the locked
All
symbol is shown, this setting can only be adjusted
by your provider.
This is a prescription setting which controls the
CPAP
operating pressure during the delay time.
Mode
The standard tubing length supplied with the device
is 6 feet (1.8 meters). However, you can adjust your
SleepCube to accommodate an air supply tubing
All
length of 10 feet (3.0 meters). Replacement tubing
can be obtained through your home care provider
(part numbers listed at the end of this manual).
Pressing the down key enters the Enable menu.
All
Refer to next section for Enable Menu details.
This setting changes the contrast for the LCD
display. This only needs to be adjusted if you find it
All
difficult to see the display from your viewing angle.
The total number of hours (in 1/10 hour
increments) that the SleepCube has been providing
All
therapy.
The total number of hours (in 1/10 hour
increments) that the SleepCube has been in
All
operation.