sPeCifiCaTions
Size ................................................ 4.2" H x 6.5" W x 6.9" D (10.7 cm x 16.5 cm x 17.5 cm)
Weight ............................................................................................................ 2.7 lbs. (1.22 kg)
Electrical Requirements ......................................................................... 100-240V~, 50/60 Hz
Maximum Power Consumption ....................................65 watts max from AC power source
Pressure Range .....................................................................................................3-20 cmH2O
Operating Temperature Range .....................................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range .................................................... 0% to 95% RH non-condensing
Operating Atmospheric Conditions ...................................................... Sea level to 9,000 feet
Storage & Transportation Temperature Range .....................-40°F to 158°F (-40°C to +70°C)
Storage & Transportation Humidity Range ............................ 0% to 95% RH non-condensing
Maximum Limited Pressure .......................................................20 cmH2O under normal use
Sound Level (tested per ISO 17510) ............................................................................. 26 dBA
Filter Specifications
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Max Flow Rates
83 L/m @ 6.5 cmH2O (1/3 max pressure)
134 L/m @ 13 cmH2O (2/3 max pressure)
163 L/m @ 20 cmH2O (max pressure)
Pressure Accuracy ..............................................................................................± 1.0 cmH2O
Pressure Swings
(cmH2O peak to peak at 500mL tidal volume, sine wave profile)
Breaths per minute
@ 6.5 cmH2O
@ 13 cmH2O
@ 20 cmH2O
Warranty .....................................................................................................Three-year limited
Equipment classification with respect to protection from electric shock.....................Class II
Degree of protection from electric shock ............................................. Type BF Applied Part
Degree of protection against ingress of liquids ................................................................IPX1
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
Mode of operation ................................................................................................ Continuous