Problem
Inaccurate
The pipe or channel where the sensor
measurements
is installed is small or too narrow. The
ultrasonic signal bounces off the sides
of the pipe or channel.
The sensor is in direct sunlight.
The temperature of the sensor
environment has changed quickly.
The sensor is too far away from the
surface of the measurement target.
Measurement
Foam, oil, debris and surface
failure
turbulence are affecting the send and
receive signal.

Replacement parts and accessories

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Gasket, vibration isolation
Holder (level) for US9003 (in-pipe) sensor
Accessories
Description
Extension cable with connectors, 15.24 m (50 ft)
Extension cable, 82.29 m (270 ft), bare wire one end, cable only
Extension cable with bare-wire and junction box, 30.48 m (100 ft)
Extension cable with bare-wire and junction box, 82.29 m (270 ft)
Extension kit for conduit, includes: 82.29 m (270 ft) cable with bare wires and junction
box with 30.48 cm (12 in.) cable and connector to logger
Note: Order the ultrasonic sensor, dispensing gun, and gel cartridges separately.
Gel cartridges (3x) with feed tubes, for the junction box (if used) to keep water out
Gel cartridge, dispensing gun
16 English
Possible cause
W A R N I N G
Solution
Install the sensor in a larger diameter pipe or
move the sensor nearer to the water surface.
Keep the downlooking sensor a minimum of
13.34 cm (5.25 in.) from the water surface. Keep
the in-pipe sensor a minimum of 0 cm (0 in.) from
the water surface.
As an alternative, increase the minimum distance
value to a location below the false target that is
affecting measurements.
Do not let the sun shine directly on the sensor.
Make a sun-shield if an alternative location is not
available. Contact technical support for ideas for
the specific environment.
Let the sensor stabilize after large changes in the
temperature of the environment.
Install the sensor near the surface of the
measurement target.
Clean the sensor. Refer to
on page 15.
Clean the sensor
Item no.
6820
9488700
Item no.
9489000
9488100
8315200
8315201
9488000
7725600
7715300