Leakage current range:
Accuracy (0 °C to +30 °C):
Capacitance range:
Accuracy (0 °C to 30 °C):
Timer:
Hum Rejection:
Voltage range:
Accuracy (0 °C to +30 °C):
Guard Terminal:
Capacitor charging time:
Capacitor discharging time:
Maximum continuous overload:
Temperature coefficient:
0,00 nA to 999 µA
±5% ±0,2 nA at all voltages
0,00 µF to 10,0 µF (For test voltages > 200 V)
±15% ±0,03 µF
0 to 90 minutes: User selects test duration
1 mA rms per kV test voltage
2 mA rms maximum
50 to1000 V a.c. or d.c. (N.B. Display does not distinguish between V a.c. and V d.c.)
±2% ±1 V
Will guard out parallel resistances to a minimum of 250 kΩ.
When measuring 100 MΩ under these conditions, the additional error will be less
than 5%.
<2,5 seconds per µF to charge to 5 kV
<2 seconds per µF to discharge to <50 V
1 kV rms
0,2% per °C (test current >100 nA)
0,1% per °C for test voltage