3. COUNTING (PCS)
Counting mode: Press PCS to enter the counting function. The scale switches from
the weighing mode to the counting mode. By pressing this button, you can switch
from one mode to another.
4. TARE
This button is used to calibrate the scale. Place an empty container on the weighing
plate until the balance indicates a stable value. Then press „ZERO" Your balance now
displays „0". To exit the tare function, remove the container and repress „ZERO": the
scale is back to normal.
5. GREEN BACKLIGHT OF THE DISPLAY
Backlight of the display is turning off automatically if there is no load on the scale. The
scale goes into standby mode. During weighing, the backlight of the display is turned
on.

6. ENERGY SAVING FUNCTION

The scale turns off automatically after the time scheduled (2, 5 or 8 minutes of inac-
EN
tivity). This function is turned off by factory default settings. In order to turn it on, full
calibration procedure must be performed in accordance with the below instruction.
Calibration requires standard weights 5 kg and 7.5 kg / or 10 kg depending on the
scale model:
1.
Turn OFF the scale.
2.
Press TARE and ON/OFF simultaneously for about 5 seconds. Display will show
SET and then oFF (FOO).
3.
Press ON/OFF – display will show 0(F01).
4.
Press ON/OFF – display will show 1.0(F02).
5.
Press ON/OFF – display will show 0.5(F03).
6.
Press ON/OFF – display will show 1.0(F04).
7.
Press ON/OFF – display will show X(F05) – where X can be 0 (AUTO-OFF is
off) or 2 (scale will turn off after 2 minutes of inactivity) or 5 (scale will turn off
after 5 minutes of inactivity) or 8 (scale will turn off after 8 minutes of inactivity).
X can be changed by pressing MODE button.
10
8.
Press ON/OFF – display will show 1(F06).
9.
Press ON/OFF – display will show 4(F07).
10. Press ON/OFF – display will show 75000(CAP) or 100000(CAP) depending on
the scale model.
11. Press ON/OFF – display will show 1(div).
12. Press ON/OFF – display will show 0.1(dEC).
13. Press ON/OFF – display will show on(LCP).
14. Press ON/OFF – display will show 5000.0(WE1).
15. Press ON/OFF – display will show 7500.0(WE2) or 10000.0 (WE2) depending on
the scale model.
16. Press ON/OFF – display will show 30......(SCL).
17. Press ON/OFF – display will show LoAd and blinking 50000 (WE1).
18. Put on the scale standard weight 5 kg – display will show some numbers and
then PASS and the weight value.
19. Then the display will show LoAd and blinking 75000 (WE2) or LoAd and blinking
100000 (WE2) depending on the scale model.
20. Put on the scale standard weight 7,5 kg or 10 kg depending on the scale model –
display will show some numbers and then PASS and the weight value.
21. End of calibration.
III. USE
Press the ON / OFF to turn on the scale. After turning on the scale, display first
shows "000000" to „999999", and then "0.0g".Let the scale to acclimate for 15min to
achieve optimum precision results.
OVERLOAD
When the weight of the object exceeds the capacity of the scale, it displays "HHHH".
Remove immediately the weight of the scale to prevent damages.
TARE
Place an empty container on the weighing plate until the balance indicates a stable
value. Then press „ZERO" Your balance now displays „0". To exit the tare function,
remove the container and repress „ZERO". The scale is back to normal.