Medela Invia Liberty Mode D'emploi page 21

Masquer les pouces Voir aussi pour Invia Liberty:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Recharge the battery either by placing the Invia Liberty pump
in the docking station or plug in the charger to the electrical outlet port
on the pump.
Remaining time of battery is approximately 30 minutes.
If the pump is in Standby mode for more than 5 minutes, an alarm will
go off.
To continue therapy press
Cool Invia Liberty pump down as per instructions on the display.
Dressing:
– Check dressing for air leakage. Press firmly around the edges of
the dressing, around the drain tubing or on the Invia FitPad.
– May apply additional film dressing to seal the leaking area.
Connectors:
– Ensure that the tubing connected to the dressing is connected properly
to the canister tubing.
– Ensure that the canister tubing is inserted straight into the pump.
Canister:
– Ensure that the canister is properly inserted, release the canister
and reposition.
– Ensure that the orange O-ring / gasket, placed beside the canister tubing
on the pump is not missing. Additional O-ring is available via Medela
representative.
Call your healthcare professional.
Remarks / potential cause of fault
"On"
[
]
21

Publicité

Table des Matières
loading

Table des Matières