Medela Invia Liberty Mode D'emploi page 13

Masquer les pouces Voir aussi pour Invia Liberty:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

WARNING
NPWT must run 24 hours a day. Do not stop the therapy
unless instructed by your healthcare professional or if
excessive bleeding occurs.
WARNING
The dressing should be changed and therapy restarted by
the healthcare professional if therapy is discontinued for
more than 2 hours.
WARNING
Invia Liberty pump does not detect blockage along tubing
connected to the drain adapter or connected to the curved
arm of the Y-connector. Check dressing tubing connected
to the drain adapter or the curved arm of the Y-connector
frequently throughout the day to make sure tubing is not
kinked or blocked.
If blockage is identified along tubing connected to the drain adap-
ter or tubing connected to the curved arm of the Y-connector,
contact your healthcare professional.
Turn OFF
Switch Invia Liberty pump OFF by pressing
[
] for 3 seconds. Press only once.
13

Publicité

Table des Matières
loading

Table des Matières