Safety Warnings and Precautions must be read and understood
before the instrument is used. They must be observed during use.
•
The circuit under test must be de-energized and isolated before
connections are made except for voltage measurement.
•
Circuit connections must not be touched during a test.
•
After insulation tests, capacitive circuits must be allowed to discharge
before disconnecting the test leads.
•
The Live Circuit Warning and Automatic Discharge are additional safety
features and should not be regarded as a substitute for normal safe
working practice.
•
Replacement fuses must be of the correct type and rating.
•
Test leads, including crocodile clips, must be in good order, clean and
have no broken or cracked insulation.
•
U.K. Safety Authorities recommend the use of fused test leads when
measuring voltage on high energy systems.
T H I S I N S T R U M E N T M U S T O N L Y B E U S E D B Y S U I T A B L Y T R A I N E D A N D C O M P E T E N T P E R S O N S
SAFETY WARNINGS
NOTE
.
2