Detailed Accuracy Specifications; Maintenance And Repair - Amprobe AMB-50 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Detailed Accuracy Specifications

Accuracy: ±([a% of reading] + [number of least significant digits]), guarantee for 1 year.
Operating temperature: 18°C ~28°C
Relative humidity: 45~75%RH
A. Voltage Measurement
Measurement Range
Resolution
Accuracy
B. Insulation Resistance Measurement
Output
500V
Voltage
Display
0.5M
e
Range
Open Circuit
DC 500V 0%~+ 20%
Voltage
Test Current
1mA~1.2mA @ 500k
0.50M ~99.9M :
100M ~9.99G :
10.0G ~20.0G :
Accuracy
Short Circuit
Less than 2.0mA
�OPERATING CAUTION
At any output voltage, when the tested resistance is less than 10M Ohm, the measurement shall
be limited to 10 seconds maximum.

MAINTENANCE AND REPAIR

If there appears to be a malfunction during the operation of the meter, the following steps should
be performed in order to isolate the cause of the problem.
1. Check the battery. Replace the battery immediately when the symbol "
LCD.
2. Review the operating instructions for possible mistakes in operating procedure.
Except for the replacement of the battery, repair of the meter should be performed only by a
Factory Authorized Service Center or by other qualified instrument service personnel. The front
panel and case can be cleaned with a mild solution of detergent and water.Apply sparingly with a
soft cloth and allow to dry completely before using. Do not use aromatic hydrocarbons, Gasoline or
chlorinated solvents for cleaning.
• Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or
solvents.
• To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals
can affect readings.
DC Voltage
±30 ~ ±600V
±(2%+5), when voltage between 30-100V (5/60Hz) the accuracy is ±(2%+8)
1000V
~20G
2M
~40G
e
e
DC1000V 0%~+ 20%
1mA~1.2mA @ 1M
e
±
(3%+5)
2.0M ~99.9M :
±
(5%+5)
100M ~9.99G :
±
(10%+5)
10.0G ~40.0G :
1V
2500V
5M
~100G
e
e
DC 2500V 0%~+ 20%
1mA~1.2mA @ 2.5M
e
±
(3%+5)
5.0M ~99.9M :
±
(5%+5)
100M ~9.99G :
±
(10%+5)
10.0G ~100G :
11
AC Voltage
30V~600V (50/60Hz)
5000V
100M
~1000G
e
e
DC5000V 0%~+ 20%
1mA~1.2mA @ 5M
e
±
10.0M ~29.9M : (For
(3%+5)
reference only)
±
(5%+5)
30.0M ~99.9M :
±
(10%+5)
100M ~9.99G :
10.0G ~99.9G :
Above 100G : [
Humidity:Below 50%]
" appears on the
e
e
±
(3%+5)
±
(5%+5)
±
(10%+5)
±
(20%+5)

Publicité

Table des Matières
loading

Table des Matières