5.7.1 Procedure when adjusting (non-verifiable models):

Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
PFB-BA-e-1630
Operation
 In weighing mode press
„UnLoAd" appears.
When display „LoAd" appears, put the necessary
adjustment weight (see chap.1; Technical Data) on the
center of the weighing plate.
Adjustment will be carried out automatically after dead
stop control.
 Wait for stability display
„Pass" appears, the adjustment process has been
finished successfully.
 Take away adjustment weight
Wait until the balance is again in the weighing mode.
approx. 3 sec. until