5.8.2 Linearization of non verifiable models (low resolution) and verifiable
models
Balance display
In case of verifiable models press the adjustment switch on the lower side of the
balance!
18
Operation
Start balance by pressing
carries out a self test, press
appears
Press
repeatedly until "tECH" appears
Press
, „Pin" appears
Press
,
appears
Press
anew, „Pin" appears again
Press successively
appears; after the apparition of the stability display
and the signal sound, „LoAd 1" will be displayed
Put on weight 1 (see table chap.6.8.1)
Wait for the stability display and the signal sound
„LoAd 2" appears
Remove weight 1 and put on the weight 2
After the stability display and the signal sound „LoAd
3" appears; remove weight 2 and put on weight 3
After the stability display and the signal sound the
balance carries out a selftest. Remove the weight
during selftest. The balance changes into weighing
mode. Now the linearization is concluded
successfully.
O N
O F F
. While the balance
until „F1 Unt"
and
successively, „P1 Lin"
,
and
PFB-BA-e-1630
, „LoAd 0"