5.7.2 Adjustment procedure (models PFB-M):
Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
Press adjustment switch on the lower side of the balance!
14
Operation
Start balance by pressing
selftest press
„F1 Unt" is displayed.
Press
repeatedly until „tECH" is
displayed
Press
, „Pin" is displayed
Enter password:
,
Press
„P1 Lin" is displayed
Press
, „P2 CAL" will be displayed
Press
, „UnLoAd" followed by „LoAd" is
displayed.
Put the necessary adjustment weight (see
chap.1; Technical Data) on the centre of the
weighing plate.
Adjustment will be carried out automatically after dead
stop control.
Wait for stability display
During the
.
,
subsequently.