5.7.2 Adjustment procedure (models PFB-M):

Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
 Press adjustment switch on the lower side of the balance!
14
Operation
 Start balance by pressing
selftest press
„F1 Unt" is displayed.
Press
repeatedly until „tECH" is
displayed
 Press
, „Pin" is displayed
 Enter password:
,
Press
„P1 Lin" is displayed
 Press
, „P2 CAL" will be displayed
 Press
, „UnLoAd" followed by „LoAd" is
displayed.
 Put the necessary adjustment weight (see
chap.1; Technical Data) on the centre of the
weighing plate.
Adjustment will be carried out automatically after dead
stop control.
 Wait for stability display
During the
.
,
subsequently.