10. Specifications
Model
Display
Cuff pressure range
Blood pressure
measurement range
Pulse measurement
range
Accuracy
Inflation
Deflation
Measurement method Oscillometric method
Transmission method Bluetooth
Wireless
communication
Operation mode
IP classification
Rating
Power source
Battery life
Durable period
(Service life)
Operating conditions
Storage / Transport
conditions
BP7455CAN
LCD digital display
0 to 299 mmHg
SYS: 60 to 260 mmHg
DIA: 40 to 215 mmHg
40 to 180 beats / min.
Pressure: ±3 mmHg
Pulse: ±5 % of display reading
Automatic by electric pump
Automatic pressure release valve
®
Low Energy
Frequency range: 2.4 GHz
(2400 - 2483.5 MHz)
Modulation: GFSK
Effective radiated power: < 20 dBm
Continuous operation
Monitor: IP20
AC adapter: IP21 or IP22
DC6 V 4.0 W
4 "AA" batteries 1.5 V or AC adapter (INPUT
AC100 - 240 V 50 - 60 Hz 0.12 - 0.065 A)
Approximately 700 measurements (using
new alkaline batteries)
Monitor: 5 years
Arm cuff: 5 years
AC adapter: 5 years
+50 °F to +104 °F (+10 °C to +40 °C) / 15 to
90 % RH (non-condensing) / 800 to
1060 hPa
-4 °F to +140 °F (-20 °C to +60 °C) / 10 to
90 % RH (non-condensing)
HEM-7342T1-ACACD6
EN
47