7.2 Models with internal weight (KERN ALJ/PLJ)

For non verified balances four adjustment possibilities are available in the menu.
Menu settings
In weighing mode press the MENU button and keep it pressed until
the acoustic signal gets mute. The first menu item „units" is displayed.
Press MENU button repeatedly until „calib" appears and confirm using
PRINT button.
Use the MENU key to choose between the following settings:
AUT-CAL
I-CAL
E-CAL
TEC-CAL
Take over selection using the PRINT button.
The balance returns to menu.
To finish the menu press the MENU button and keep it pressed until
the acoustic signal gets mute. The balance returns automatically into
weighing mode.
AUT-CAL
With activated AUT-CAL function the internal adjustment is
Factory setting of
automatically started when the balance
verified balances
The automatic adjustment function is always enabled. You can
start adjustment at any time by pressing the CAL-key manually.
32
Automatic adjustment with internal weight.
Factory settings of models with type approval.
Adjustment with internal weight after having pressed
CAL; not available for models with settings appropriate
for verification.
Adjustment with external weight
not available for models with settings appropriate for
verification.
not documented
• after the weighing balance was disconnected from
the mains
• after pressing ON/OFF in stand-by mode
• after a temperature change of 1.5 °C
with non loaded weighing plate / zero display
• after a time interval of
with non loaded weighing plate / zero display
3 hours