TECNO-GAZ
10. ALARM MESSAGES
If a fault arises do not attempt to resolve it yourself, ask for the assistance of an authorised technician.
Er 0
10.0
: POWER FAILURE
Er 2
10.2
: DISINFECTION INCOMPLETE
This occurs when the temperature does not reach the required value for disinfection.
Er 3
10.3
: DOOR LOCK
This condition happens when, during a cycle, the reel is disactivated, so the door is unblocked. Or, after the start the reel do not activate
its limit swicth within the time setted by the parameter.
Er 4
10.4
: DOOR OPEN
This occurs when the limit switch of the door lock is disengaged during a work cycle and the door results as locked but open
Er 5
10.5
: WATER SHORTAGE
This occurs when solenoid valve is open but the tank does not reach the required water level within the set time. Possible causes for
this condition are lack of water or incorrect parameter setting.
Er 6
10.6
: HEATING ELEMENTS PROTECTION
This condition happens when,with heaters ON, the disinfection do not ends within the time setted by the parameter. Likely reasons of
such a block could be either a problem on eating elements or a too low setted time by the parameter.
Er 7
10.7
: TEMPERATURE PROBE 1
This occurs when the bath temperature probe is damaged.
Er 8
10.8
: TEMPERATURE PROBES DIFFERENT
This occurs when the temperature probes is different.
Er 9
10.9
: TEMPERATURE PROBE 2
This occurs when the bath temperature probe is damaged.
Er 10
10.10
: DISCHARGE PUMP
This occurs when the presence of liquid is detected following the discharge stage.
In case of short voltage drop the machine can give ER3(door lock) instead of POWER FAILURE: if when alarm ER3 appears the
door is unlocked, and consequently it can be opened, the error is without doubt due to a short no power.
N.B.: The power failure error is given only when there is no power during a working cycle.
WARNING:
ASSISTANCE:
maintenance, contact our service assistance centre, describe the fault and give the
model and serial number of the machine.
This occurs when there is an electrical power failure during a work c
Use specific products to remove scale, avoid using highly corrosive
products.
If your machine is not working properly even after routine
32