3. USING THE MACHINE

3.1 Checks
Check the quantity of chemical additives present and top-up if necessary.

3.2 Opening and closing the door

The machine is fitted with an electric door lock to prevent it being opened when the machine is running.
To open the door during a wash cycle, interrupt the cycle and remember that:
3.
The items inside the machine may be very hot.
4.
The entire wash cycle must be repeated.

3.3 Preparation

Place the items to be washed inside the machine and position them carefully on the holder and in the rack.
Items should not overlap.
Receptacles should be positioned so that liquids can flow out easily.
Tall or heavy items should be placed towards the middle of the rack if possible to facilitate washing.
Make sure that nothing is blocking the arms and that they turn freely.
Caution: the maximum load for each cycle is 8.0 kg.
3.4 FUNCTIONING
Place the items to be washed in the relative tray positioning them carefully and follow the instruction that follows. The control panel with
display is illustrated in the diagram.
This panel makes the machine easy to use as it indicates the stage of the cycle in progress, the maximum temperature reached during
disinfection and fault messages.
3.5 Treatment of turbines and straight and angular hand pieces (opposed-angles)
Your machine can be equipped with a special basket studied specifically for washing hollow instruments, that need to be washed and for
thermodisinfection both inside and outside the cavity.
CODE: C80
The above-mentioned basket is equipped with special accessories suitable for inserting the hand pieces. These supports are made in
two parts, screwed together, between which there is a special filter.
There is a rubber adaptor in the top part, available in two different diameters for the hand pieces.
We recommend washing the filters in the fitting for supporting the turbines and the hand pieces on a weekly basis, or replacing them
with new filters.
The following operations should be performed for the correct treatment of the turbines and straight and angular hand pieces:
-
prewash with cold water, to eliminate residues of blood and saliva.
wash at 45°C, adding mineral-free, neutral-PH, liquid detergent.
-
-
thermodisinfection at a temperature of 90°C for 1 minute, adding the additive for eliminating the residual water.
The following chemical products are recommended:
-
MEDICLEAN neutral-PH detergent.
-
MEDIKLAR, additive for eliminating the residual water.
PRECAUTIONS
Micromotors cannot be subjected to thermodisinfection treatment.
The thermodisinfection treatment cannot be done with the 90°C program for 3 or 10 minutes.
Never use powder detergents.
(for the user)
25
DAQF028_rev0.10