10.2 Errors
In case of error alarm the device has to be reset to start a new cycle.
Display
22
ALARM RESET
Cause
Water infeed alarm
- the tap is closed
- filters clogged
- pressure of the water
system not normal
Water level alarm
- the device has been
moved or tilted during the
cleaning.
- too many instruments have
been added after water
infeed.
Water drain alarm
- filters clogged
- drain hose in wrong
position.
- Wrong installation: anti-
return valve on the drain
function
Keep pressed for 8s. button
OK
Solution
- Make sure the tap is open.
- Clean tank filter, drain filter, and water
infeed filter.
- Make sure water pressure ranges are
within the indicated values. If the
pressure is insufficient, contact a
plumber.
-
Make
sure
the
(OPTIONAL) is not locked.
- Make sure there is no water into the
tank and start a new cycle
- Make sure there is no water into the
tank and start a new cycle.
- Do not add instruments during the
cycle.
- Clean the tank filter and the drain
filter.
- Position the drain hose correctly, see
installation.
- Make sure there is no water into the
tank and start a new cycle.
Water
Block