Begin the procedure following standard hospital guidelines. Throughout the injec on
SAFIRA is designed to limit the infusion pressure up to a maximum of 20 psi.
12.
Should an infusion stoppage be encountered during the procedure, it may indicate an
intraneural injec on or a blockage in the fluid path. Verify needle patency and/or
reposi on needle to allow low pressure infusion and then reset the Flashing Red 
Light/Infusion Stopped on the Driver.
13.
To reset the Flashing Red Light / Infusion Stopped on the Driver, follow either of the
two (2) steps indicated below:
a. Aspirate (Yellow side closest to the cable of the Foot Pedal) un l red lights go
out: or
b. Release Sterile Syringe (see Diagram #5) and re‐click into place
Now the procedure can be con nued.
14.
When the procedure is complete, follow standard hospital protocol for the following:
a. Dispose of the Sterile Syringe using acceptable standard prac ce for biohazard
waste
b. Driver and Foot Pedal components should be stored according to hospital
prac ce for reuse
15.
End of Life: Once the end of life indicator on the Driver (steady red lights) comes on,
the component should be disposed of in accordance with hospital procedure. Do not
start a procedure if the end of life indicator is already on from the beginning. If the
end of life indicator lights up during a procedure, immediately disconnect the syringe
and either replace the Driver or complete the procedure manually.
 
Disposal:
For SAFIRA components, follow standard hospital protocol for disposal.
1. 
Sterile Syringe:  Use acceptable prac ce for biohazard waste. 
2. 
Driver: Should not be disposed of as unsorted municipal waste. Dispose of unit in line
with local guidelines. 
3. 
Foot Pedal: Should not be disposed of as unsorted municipal waste. Dispose of unit
in line with local guidelines. 
MP‐0007‐0015 ver.4.0
Page 8 of 48