10.
Driver Indicator Lights: On the top edge of the Driver (refer to Diagram #7) colour
indicator lights are displayed represen ng various func ons in ac on or func ons
that require addi onal ac on. The indicators are as follows:
LED Colour
State
Red
Flashing
(Infusion has
stopped)
Red
Steady
(End of Life)
Green
Steady
(Infusion Func on
Ac ve)
Yellow
Steady
(Aspira on Func on
Ac ve)
No Ac ve
N/A
Light
Green =
Yellow =
Infusion
Aspira on
Diagram #7: Top edge of driver, illustrated with all indicators.
MP‐0007‐0015 ver.4.0
Meaning
Warning:
Operator: Infusion has stopped;
Immediate
1. Verify needle patency and/or
response by
reposi on needle to allow low
operator is
required
2. Reset SAFIRA (see steps 12 and
Warning:
Operator: Immediately disconnect
Immediate
the syringe and either replace
response by
Driver or con nue the procedure
operator is
required.
Normal: System
Operator: Con nue infusion
opera ng as
ac vity un l medical result is
designed indica ng
infusion ac vity
Cau on: Aspira on
occurring ‐ system
opera ng as
intended
Normal: System is
ready for infusion
or aspira on
Flashing Red =
Steady Red =
Infusion Stopped
End of Life
Operator Response
pressure infusion
13 for details)
manually.
achieved.
Operator: Follow Standard
Precau ons during use of
aspira on func on
Operator: N/A
Page 7 of 48
GB