Electro Magne c Immunity: (con nued)
Guidance and Manufacture's Declara on: Electromagne c Immunity
SAFIRA is intended for use in the electromagne c environment specified below. The
customer and/or user of SAFIRA should assure that it is used in such an environment.
Immunity
Test
Conducted
RF IEC 61000
‐4‐6
6V rms ISM and
Radiated RF
IEC 61000‐4‐
3
Note 1: The guidelines may not apply in all situa ons, Electromagne c propaga on is
affected by absorp on and reflec on from structures, objects and people.
MP‐0007‐0015 ver.4.0
IEC 60601 Test
Compliance
Level
Level
3V rms (1kHz
3Vms
80%) 150kHz –
80MHz
6V rms
amateur radio
bands
3V/m (1kHz
3V/m
80%) 80MHz –
2.7GHz
Electromagne c Environment: Guidance
Portable and mobile RF communica on
equipment should be no closer to any part of
SAFIRA, including cables, than the
recommended separa on distance of 300mm.
Field strength from fixed RF transmi ers, as
determined by an electromagne c site
a
survey
, should be less than the compliance
level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
b
.
Page 12 of 48