Electro Magne c Emission: 
Guidance and Manufacture's Declara on: Electromagne c Emission 
SAFIRA is intended for use in the electromagne c environment specified below. The
customer and/or user of SAFIRA should assure that it is used in such an environment.
Emission Test 
Compliance 
RF Emissions 
CISPR 11 
RF Emissions 
CISPR 11 
Guidance and Manufacture's Declara on: Electromagne c Immunity 
SAFIRA is intended for use in the electromagne c environment specified below. The
customer and/or user of SAFIRA should assure that it is used in such an environment.
Immunity Test 
IEC 60601 Test 
Electrosta c
±8 kV contact
Discharge (ESD)
±2, 4, 8, 15 kV
IEC 61000‐4‐2
Power Frequency
30 A/m
(50Hz/60Hz)
Magne c Field
IEC 61000‐4‐8
Note: U
is the a.c. mains voltage prior to applica on of the test level.
T
MP‐0007‐0015 ver.4.0
Electromagne c Environment: Guidance 
Group 1 
SAFIRA uses RF energy only for its internal func on.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
Group A 
SAFIRA is suitable for use in all establishments, other
than domes c and those directly connected to the
public low‐voltage power supply network that
supplies buildings used for domes c purposes.
Compliance 
Level 
Level 
±8 kV contact
±2, 4, 8, 15 kV
air
air
30 A/m
equipment.
Electromagne c Environment: Guid-
ance 
Floors should be wood, concrete or
ceramic le. If floors are covered with
synthe c material, the rela ve
humidity should be at least 30%.
Power frequency magne c fields
should be at levels characteris c of a
typical loca on in a typical
commercial or hospital environment.
Page 11 of 48