2.
Open the sterile package and remove Sterile Syringe. Use normal hospital technique
to fill and prepare the syringe.
3.
Connect the appropriate needle to the luer fi ng of the syringe. Ensure that the
needle is connected securely to the syringe. Once secured, prime the needle and
tubing set to remove the air.
Diagram #1: Needle prepared to connect to luer fi ng of SAFIRA syringe
4.
To a ach the Sterile Syringe to the Driver; remove protec ve foam cap from the
Driver gear prior to connec ng (see below). Align Sterile Syringe with Driver as
shown in Diagram #2. Once aligned, gently press the Sterile Syringe in the Driver. It
will be seated correctly when you hear it "click" into place.
Diagram #2: SAFIRA ready for the Sterile Syringe
5.
A ach the Foot Pedal to the Driver. The plug end of the cord goes into the round
receptacle opening as shown in Diagram #3. Place Foot Pedal on the floor.
6.
Press the infuse part (green end, furthest from the cable) of the Foot Pedal to check
that the tubing is correctly primed and the SAFIRA system is opera ng as expected.
MP‐0007‐0015 ver.4.0
Diagram #3: Foot Pedal plugged in
GB
Page 5 of 48