7.
Place the Foot Pedal on the floor below the surgical site in a convenient loca on for
the Anaesthesiologist. See Diagram #4 for reference.
8.
Removal: In order to remove the Sterile Syringe from the Driver, simply press the
square teal bu on on the side of the Driver as shown in Diagram #5. The Sterile
Syringe will immediately release from the Driver.
9.
The Foot Pedal has two colours: Green = Infusion; and, Yellow = Aspira on as shown
in Diagram #6.
The "aspira on 
func on" is
accessed by
stepping on the
yellow half of
,
the Foot Pedal
the end closest 
to the cable.
Diagram #6:  Foot Pedal showing Infusion and Aspira on func ons
MP‐0007‐0015 ver.4.0
Diagram #4:  Foot Pedal placed on floor below surgical site
Diagram #5: Press square release bu on to remove Syringe
Press square
teal release
bu on to
remove
Sterile
Syringe
The "infusion 
func on" is
accessed by
stepping on the
green half of the
Foot Pedal,
furthest from 
the cable.