Electrocution and fire hazard. Attach the power cord so that it cannot touch hot surfaces.
Electrical shock and fire hazards. Replacement power cords must:
• Have the correct plug style for the outlet connection
• Have a rating sufficient for the supply voltage and current. Refer to the requirements in the
Specifications section.
• Meet or exceed local electrical code requirements
Personal injury hazard. Instruments or components are heavy. Use assistance to install or move.
Make sure that there is at least 15 cm (6 in.) of clearance around the instrument to prevent heat damage to
objects near the instrument and so there is sufficient air flow around the instrument.
This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Use of this instrument at an
altitude higher than 2000 m can slightly increase the potential for the insulation to breakdown, which
can result in an electric shock hazard. The manufacturer recommends that users with concerns
contact technical support.
1. Put the instrument on a stable, level, heat-resistant surface. Do not put the instrument in direct
sunlight.
2. Connect the power cord to the power connector. Refer to
3. Connect the power cord to an electrical outlet with protective earth ground.