5
Technical Information
PIN Functions
Old IPCU Releases - Compatibility Report Type 6
Report Format: "HVc sssss ff p iiiiixA vvvvV <CR>"
Report Fields:
HVc
sssss
ff
p
iiiii
xA
vvvv
Example:
RT 6
+
HV2 OFF 00 0
HV1 OFF 00 0
HV2 OFF 00 0
HV1 OFF 00 0
HV2 OFF 00 0
96/104
HV Channel c= 1 or 2
HV Channel Status ['OFF__', 'ON___', 'FAULT']
8 bits HEX representation of HV Channel Faults
'0'=Start, '1'=Protect
HV Channel Current in xA
nA or uA (nano Ampere or micro Ampere)
HV Channel Voltage in V (Volts)
0uA
0V
0uA
0V
0uA
0V
0uA
0V
0uA
0V
Ion Pumps Control Unit User Manual / 87-900-124-01
(C)