Télécharger Imprimer la page

Agilent Technologies 629-5001 Notice De Mode D'emploi page 76

Publicité

5
Technical Information
ION Pump Controller Description
Output for the pumps:
With Line Supply:
Output Voltage (each channel)
Output Current (each channel)
Maximum Output Power (channel 1)
Maximum Output Power (channel 2)
Operating Temperature
Storage Temperature
Compliance to Norms
Line Cable
Line fuse
High Voltage Connection
Battery Connection
(*) HV Battery Back-Up Option if available for the model.
All connecting cables for remote I/O and serial connections, if longer than 3 m,
NOTE
must be of the shielded type.
76/104
Default value =+5000V +1/-5 %, settable
from +3000 to +5000Vdc at steps of +/-
50V by Serial Line command
30 mA ± 10 % for 40 W output, 60 mA +/-
10 % for 80 W output
40 W
80 W
5 °C to 40 °C
-20 °C to 70 °C
Radio Interference: EN55011 Class A
Group 1
Harmonics:
Flickers:
ESD:
Burst:
Radiated RF Immunity:
Surge:
Conducted noise imm.:
Line supply interruptions: EN61000/4/11
Safety:
3 m long cable (3 wires), no plug
Two T5A
2 High Voltage Fischer connectors
3 pin connector (for Lead Acid Battery 12
Vdc 24 Ah)
Ion Pumps Control Unit User Manual / 87-900-124-01
EN61000/3/2
EN61000/3/3
EN61000/4/2
EN61000/4/4
EN61000/4/3
EN61000/4/5
EN61000/4/6
EN61010-1
(C)

Publicité