5
Technical Information
PIN Functions
User Operator Report Type 2 and User Technician
Report Type 3
Report Format:
"HVc sssss iiiiixA vvvvV F=ffff E=eeee pppppmW i=yyyy v=zzz
c=www <CR>"
Report Fields:
HVc
sssss
iiiii
vvvv
ffff
eeee
ppppp HV Channel Power in mW
yyyy
zzz
www
Example:
RT 2[Enter]
+
HV2 OFF
HV1 OFF
HV2 OFF
94/104
HV Channel c= 1 or 2
HV Channel Status ['OFF__', 'ON___', 'FAULT']
HV Channel Current in nA or uA
HV Channel Voltage in V
Hexadecimal representation of HV Channel Faults
Hexadecimal representation of HV Channel Events
16 Bits hexadecimal Current
10 Bits hexadecimal Voltage
12 Bits hexadecimal Control Output
0uA
0V F=0000 E=0000
0uA
0V F=0000 E=0000
0uA
0V F=0000 E=0000
Ion Pumps Control Unit User Manual / 87-900-124-01
0mW i=0001 v=000 c=D80
0mW i=0003 v=000 c=D80
0mW i=0001 v=000 c=D80