Technical Information
PIN Functions

Controller Commands Mode

HV Channel Status Setting
"A0ns" where :
HV Channel Start/Protect Mode Setting
"C0ns"[Enter] where :
Clear Fault (Reset)
"F0n"[Enter] where:
Tab. 6
Err #
[0001]
[0002]
[0004]
[0008]
[0010]
[0020]
[0040]
[0080]
[0100]
86/104
A0 = HV Channel Serial Command for Status Set
N
= HV Channel Number 1, or 2
s
= 0 for OFF, 1 for ON
C0 = HV Channel Serial Command for Start/Protect Setting
N
= HV Channel Number 1 or 2
S
= 0 for START mode, 1 for PROTECT mode
F0 = HV Channel Serial Command for Fault reset
n
= HV Channel number 1 or 2
Bit positional Fault table when HV ON
Description
Cable Interlock when HV On
Remote Interlock when HV On
HV Over Voltage when HV On
Bad Current Offset starting HV On
HV Over Power when HV On
Panel Switch Moved when HV On
HV Over Temperature when HV On
HV Over Current – Protect=Enable
HV Under Voltage when HV On
Ion Pumps Control Unit User Manual / 87-900-124-01
Front panel Led
1 blink of 200 ms every 4 s
2 blink of 200 ms every 4 s
3 blink of 200 ms every 4 s
4 blink of 200 ms every 4 s
5 blink of 200 ms every 4 s
6 blink of 200 ms every 4 s
7 blink of 200 ms every 4 s
8 blink of 200 ms every 4 s
9 blink of 200 ms every 4 s