Tab. 7
Err #
[0001]
[0002]
[0004]
[0008]
[0010]
[0020]
[0040]
[0080]
[0100]
[0200]
Max Output Voltage setting
"H0nxxxx"[Enter] where:
Set-Point Current Threshold Setting
(Minimum Access Level Required: "0")
"WR[space]Y[space] xxxxxx"[Enter]
Ion Pumps Control Unit User Manual / 87-900-124-01
Bit positional Event table when HV On or OFF
Description
Cable Interlock when HV Off
Remote Interlock when HV Off
HV Over Voltage when HV On
Bad Current Offset when HV Off
HV Over Power when HV On
Spare
Over Temperature
HV Over Current – Protect=Disable
HV Under Voltage when HV On
Current Below Threshold when HV On
H0 = HV Channel Serial Command for Voltage Target
n
= HV Channel number 1 or 2
xxxx = Voltage Target (min 3000V, max 5000V in 50 V steps)
WR = Write
Y
= 14 for channel 1
= 15 for channel 2
xxxxxx = Current Set-Point value in nA (from 0 to 100000000)
Technical Information
Front panel Led
No blink
No blink
No blink
No blink
No blink
No blink.
No blink
No blink
No blink
No blink
(C)
5
PIN Functions
87/104