AIR COOLING KIT INSTALLATION
TV 301SF with Navigator Controller
An air cooling kit (mod. 969-9299) is available to
improve the TV 301SF cooling during heavy op-
erational conditions (optional).
Fan specifications:
3
− air flow: 140 m
/h
− input voltage: 24 Vdc
− dimensions: 119 x 119 x 32 mm
− power: 2.6 W
The fan bracket is shaped so that it can be
mounted close to the pump.
When the pump is supplied by the rack C.U. 969-
9437 or 969-9537 please utilize the special fan ex-
tension cable 969-9949.
To fix the fan to the TV 301SF case execute the
following procedure (see the following figure):
1. If the controller will be installed on board at the
bottom of the pump, fix the fan bracket to the
pump body between the pump and the control-
ler by means of the fixing studs (the same
used to fix the controller on board);
2. Connect the fan supply to the P4 connector of
the controller.
When the controller will be side-mounted:
1. Fix the fan bracket by means of the furnished
screws;
2. Connect the fan supply to the P4 connector of
the controller.
82/89