15.3 Specifications
15.3.1 Accuracy
■
Specified accuracy ± (% of the reading + display error in counts).
■
Accuracy is maintained for 1 year at +23 °C (± 5 °C), ≤75 % RH (non-con-
densing).
■
Temperature coefficient: +0.1 x (specified accuracy) / 1 °C.
■
To avoid impaired measurements, do not operate the product in areas with
electromagnetic field strengths >1 V/m (±5 %).
15.3.2 Calibration
■
The recommended calibration interval is 1 year.
■
Calibration should only be performed by qualified personnel.
15.3.3 Alternating current (AC)
Range
4.000 A
40.00 A
*Accuracy: 5 - 100 % of the measuring range
■
Overload protection: 600 V 40 A
■
Frequency range: 50-60 Hz
■
Permitted display with unused measuring input: ≤5 counts (clamp closed)
True RMS Crest factor (CF) for non-sinusoidal signals: max. 3.0:
■
Crest factor 1.0 – 2.0 : +4 % deviation
■
Crest factor 2.0 – 2.5 : +5 % deviation
■
Crest factor 2.5 – 3.0 : +7 % deviation
Resolution
0.001 A
0.01 A
Accuracy*
±(4.0 % +10)
with a low-pass filter
(VFC) ±(6.0 % + 20)
±(4.0 % +9)
with a low-pass filter
(VFC) ±(6.0 % + 20)
65