Burkert 8221 Manuel Utilisateur page 5

Masquer les pouces Voir aussi pour 8221:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Type 8221 Hygienic Conductivity Sensor Operating Instructions
Assembly
Please check if the process connection is sufficiently clean. Then fit the conductivity sensor.
Please follow the instructions mentioned below.
Examples:
Ø 60mm min
ATTENTION
The cell constant and the linearity of the sensor can vary with the fitting situation. A
symmetrical setup is recommended. Leave an open space of 60 mm minimum diameter.
Partitions made of non-conductive material should preferably be used.
ATTENTION
A symmetrical setup is recommended in order to ensure a high degree of linearity. To
achieve high precision the cell constant should be calibrated in the final setup.
Make sure that all 4 electrodes are completely and continuously immersed in the
measuring sample.
Ø 60mm min
Page 3 of 8

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

00557719005588840055912000559121

Table des Matières