Télécharger Imprimer la page

OHAUS PX12001 Manuel D'instruction page 56

Publicité

EN-52
Download from Ohaus' Website
1. The New Hardware Wizard guides you
through the required steps to select the driver
that is located on the website.
2. After clicking Finish, the virtual port should be
ready for use.
Windows
®
typically adds the virtual port in
sequence after the highest number COM port.
For example, on PC's equipped with up to 4
COM ports, the virtual port will be COM5.
When using the USB interface with programs
that limit the number of COM port
designations (e.g. Ohaus MassTracker allows
only COM1, 2, 3, & 4), it may be necessary to
assign one of these port numbers to the new
virtual port.
This can be done in the Port Settings of the Device Manager utility, found in the Windows Control Panel.
USB INPUT
The balance will respond to various commands sent via the interface adapter.
Terminate the following commands when with a [CR] or [CRLF]
PX Commands
P
same as pressing Print
SP
print stable weight only
IP
immediate print of displayed weight (stable or unstable)
CP
Continuous print of weights
T
same as pressing Tare
Z
same as pressing Zero
PV
print software version
xT
establish a preset Tare value in displayed unit. X = preset tare value.
Sending 0T clears tare (if allowed).
Auto-Print Operation
Once Auto-Print is activated in the menu, the balance will send data as required.
If there is data in the print buffer the printer will finish printing this data.
10. SOFTWARE UPDATES
Ohaus is continuously improving its balance software. To obtain the latest release, please contact your Authorized
Ohaus Dealer or Ohaus Corporation.
Example of Windows XP Hardware Wizard
PX Series Balance

Publicité

loading