5.3.13 Auto Off
Off = disabled
30 minutes = turns off if there is no motion for 30 minutes
1 hour = turns off if there is no motion for 1 hour
2 hours = turns off if there is no motion for 2 hours
5.3.14 Capacity Bar
Off = disabled
On = enabled
When the capacity is set On, a capacity bar will display at the bottom of the screen. The capacity bar will roughly
show the current weight as a percentage of balance capacity. When the display is at zero, the capacity bar will not
display.
5.3.15 Approved Mode
Use this menu to set the Legal for Trade status.
Off = standard operation.
On = operation complies with Legal Metrology regulations.
Note: When Approved Mode is set to On, the menu settings are affected as follows:
Calibration Menu:
For InCal models, only Internal Calibration is available. All other functions are hidden.
Balance Setup Menu:
Filter Level is locked at the current setting.
Auto Zero Tracking is limited to 0.5 Division and Off. The selected setting is locked.
Auto Tare is locked at current setting.
Graduations are forced to 1 Division and the menu item is hidden.
Communication Menu (Communication->Print Settings->Print Output):
Stable Weight Only is locked On.
Numeric Value Only is locked Off.
Communication Menu (Communication->Print Settings->Auto Print):
Auto print mode selections are limited to Off, On Stability, and Interval. Continuous is not available.
Lockout Menu:
Menu is hidden
Note: The security switch located at the rear of the balance must be in the locked position to set Approved Mode to
On. The security switch must be in the unlocked position to set Approved Mode to Off. See Section 6.