PX Series Balance
To start batch weighing:
1. Re-enter the Weighing mode. Place the first sample
on the pan and press the Function / Mode button to
store the first weight.
2. Then, the message "Place sample 2" will display.
3. Remove the first sample, add the second sample to
the pan, and press the Function / Mode button to
store the second weight.
4. Continue adding components, and press the
Function / Mode button to store the weights.
To end batch weighing and print the result:
Press the Print button to end batch weighing and
print the result.
Note: a total of 10 batch of samples can be weighed.
When 10 is reached, the message "End" will display if
the Function / Mode button is pressed again. If
continue to press the button, the message "Out of
range" will display.
4.2
Parts Counting
Note: Before using any application, be sure the balance has been leveled and calibrated. The minimum piece
weight should be no less than 0.1d. In the LFT mode, the minimum piece weight is 3e, the minimum Sample Size
is 10.
Use this application to count samples of uniform weight.
Parts Counting
1. Press Tare or Zero if necessary to begin.
2. Press and hold the Function / Mode button until
Parts Counting appears.
EN-9