EN-42
Symptom / Display
8.4
Service Information
If the troubleshooting section does not resolve your problem, contact an Authorized Ohaus Service Agent. Please
visit our website www.ohaus.com to locate the Ohaus office nearest you.
Possible Cause
A wrong weighing pan is used when
power on.
The load on the pan is over the initial
weight setting.
A wrong weighing pan is used when
power on.
The pan is not installed when power on.
The weight on the pan is too heavy.
The pan is not properly installed.
System error
System error
System error
Remedy
Use OHAUS original weighing pan before
power on.
Remove the load from the pan before power
on.
Use OHAUS original weighing pan before
power on.
Install the weighing pan before power on.
Reduce sample size until the weight is
within the weighing capacity.
Install the weiging pan properly
Please contact OHAUS service
(www.ohaus.com - More - Contact Us)
Please contact OHAUS service
(www.ohaus.com - More - Contact Us)
Disconnect the power and restart.
If error persists, please contact OHAUS
service (www.ohaus.com - More - Contact
Us)
PX Series Balance