CALIBRATION (CONT)
8. When the zeroing process is complete the display will
prompt for entry of the calibration weight. Minimum
calibration is 300 lb / 136 kg. Maximum calibration
weight is 1000 lb / 454 kg. Calibration weight can be
adjusted in 10 lb or 10 kg increments. Press
the calibration weight is set.
9. The display will prompt for the calibration weights to be
placed on the platform. Press
weights are placed on the platform.
10. The display will show "Calibrating" during the calibration
process.
11. If calibration is successful, the display will show
"Calibration Complete" with a 4 digit calibration count.
If calibration is unsuccessful, the display will show an
12.
error message that the entered calibration weight and
the certified weights do not match. The display will then
revert back to the set calibration weight screen for entry
of calibration weight.
82-00014 2900KL-AM 2900KG-AM REV B