Gasket Maintenance
Periodically check the gaskets around the door for
punctures or tears. Leaks are indicated by condensation
or frost which forms at the point of gasket failure. Make
sure that the cabinet is level (refer to Location for leveling
information).
Keep the door gaskets clean and frost free by wiping gently
with a soft cloth.
To check the door seal, complete the following steps:
1. Open the door.
2. Insert a strip of paper (a couple of inches wide)
between the door gasket and the cabinet flange and
close the door.
3. Slowly pull the paper strip from the outside. You should
feel some resistance.
4. Repeat this test at 4-inch intervals around the door. If
the door does not seal properly, replace the gasket.
Alarm Battery Maintenance
Have a certified technician replace the alarm battery every
twelve months at most or when the alarm is active. The
part number for a replacement battery is 322533H02.
Preparation for Storage
If the unit is going to be stored in an off condition, allow
the unit to warm up and dry out with the door open before
moving into storage. Access the direct function and change
boF to "On".
Note: See operation section for more details about
preparing the battery for storage mode.
Probe Recalibration
Schedule
The display probe on Pharmacy models should be
calibrated every 3 years.